Ultrospec 7500 User Manual
•
41 56 2050 REV03
•
61
Step 4
Select the sample name to bring up the sample window, enter a sample
seed prefix and the incremental sample number starting value. Confirm
the settings using the confirm icon.
Step 5
Select one of the predefined units, “µg/ml”, “ng/µl”, “µg/µl”, or “mg/ml”.
Step 6
Proceed to the next parameter screen using the right/forward arrow.
Step 7
Select the dye type from one of the predefined dyes.
Each dye has fixed associated parameters:
Dye-Type
Dye Abs Max
Dye Ext Coeff
Dye Correction
Alexa Fluor 350
346 nm
19000
0.190
Alexa Fluor 405
401 nm
34000
0.700
Alexa Fluor 488
495 nm
71000
0.110
Alexa Fluor 647
650 nm
239000
0.030
Cy3
550 nm
150000
0.080
Cy5
649 nm
250000
0.050
DyLight 649
654 nm
250000
0.040
DyLight 488
493 nm
70000
0.150
FITC
495 nm
68000
0.300
Pacific Blue
416 nm
46000
0.200
r-PE
566 nm
200000
0.180
Texas Red
595 nm
80000
0.180