5
ULPA filter life: With the usage time increasing, dust and bacteria accumulate inside ULPA
filter. Filter Resistance is getting bigger, when it reaches the maximum point, there will be
audible and visual alarm. Please replace new ULPA filter, otherwise it will affect the safety
performance of the equipment. The used filter should be processed as medical waste.
There is a negative passage surrounding the work area, which is sealed strictly in the factory.
The operator is not allowed to remove or loose screws of those parts. If necessary, please
contact service personal.
Front Grille is used for air intake and drain. Do not block it, otherwise it will affect airflow.
Armrest is recommended to solve this problem and reducing the operator's wrist fatigue.
Long-term use of Microbiological Safety Cabinets will inevitably cause pollution (e.g. ULPA
filters, corner cabinets, etc.). In order to sterilize thoroughly every 500 hours, formalin
(formaldehyde) fumigation sterilizer is recommended. After sterilization, neutralize
formaldehyde gas with ammonium hydrogen carbonate. Make sure no sterilization gas
escapes during the whole process.
The maximum storage period is one year. If the period is more than one year, performance
test should be done.
In transportation process should take appropriate protective measures according to diagram
on the surface of package. Purchaser in receiving should carefully check the integrity of the
box. if there is damage and crush, please refuse to accept. And timely contact with our
company.
When Microbiological Safety Cabinet installed and used move again, please timely contact
with our company.
The ground bearing requirements: ground bearing capacity ≥350KPa.
Total weight: 145kg.
Allowable pressure of water socket: ≤0.8MPa. (Optional)
For experiments that produce waste liquid, please timely drain wastewater in the sump
through drain valve. If the liquid waste is harmful to human’s body, according to the grade of
pollution should add the corresponding grade container, or connect to the corresponding
grade drainpipes, and drain waste into the experimental equipment to immediately dispose.
Person can operate Microbiological Safety Cabinet only if he is qualified after training,
during using Microbiological Safety Cabinet if it is a harmful experiment, operator must wear
appropriate protective gloves, masks and lab coats, etc., and avoid touching mouth, eyes and