14
Picture8
1. Drain valve connecter
2. Shim (Inner diameter*outer diameter*thicknessΦ20*Φ28*2mm)
3. Safety cabinet bottom installation holes
4. Ball coupling fastening nut
5. Rubber gasket (Inner diameter*outer diameter*thicknessΦ13*Φ19*2mm)
6. Drain valve
Take out drain valve coupling, shim(Inner diameter*outer diameter*thicknessΦ20*Φ28*2mm), Ball
nut, Rubber gasket(Inner diameter*outer diameter*thicknessΦ13*Φ19*2mm), Drain valve,
assembling from up to down as Picture 8 illustrated.
Don’t place equipment where difficult to operate and disconnect device.
3.5 Checking after installation
Checking Items
Normal situation
Wind speed display
Inflow
(
0.53±0.025
)
m/s ,downflow
(
0.33±0.025
)
m/s
Blower
Blower runs normal
Fluorescent lamp
Lamp lights after pressing button
UV Lamp
Lamp lights after pressing button (please
close the glass door first)
Display screen buttons
All buttons can be used
Socket
Press the socket key, multimeter testing
output supply voltage