6.EMC
1) The device meets the emission and immunity requirements of EN 61326:2013.
2) The device is tested according to classA.
Biosafety cabinets need to take special precautions in terms of electromagnetic compatibility (EMC),
and must be installed and used according to the EMC information provided in this manual.
Port
Phenomenon
Basic standard
Test value
Perform-ance
criterion
Enclosure
Electrostatic discharge(ESD)
Electromagnetic field
Power frequency magnetic field
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-8
4kV contact discharge
8kV air discharge
3 V/m (80 MHz to 1 GHz)
3 V/m (1,4 MHz to 2 GHz)
1 V/m (2,0 MHz to 2,7 GHz)
3 A/m (50Hz,60Hz)
f
B
B
A
A
A
A
AC power
(including
protective
earth)
Voltage dip
Short interruptions
Burst
Surge
Conducted RF
IEC 61000-4-11
IEC 61000-4-11
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
0% during half cycle
0% during 1 cycle
70% during 25/30
e
cycles
0% during 250/300
e
cycles
1kV (5/50 ns, 5kHz)
0,5kV
a)
/1kV
b)
3V (150kHz to 80MHz)
B
B
C
C
B
B
A
Portable and mobile RF communication device can affect the operation of biosafety
cabinet.
In addition to the converters and cables allowed by the manufacturer of the biosafety
cabinet as replacement parts for internal components, the use of accessories, converters and
cables that are not specified may increase the radiation or reduce the anti-interference
capability of the biosafety cabinet
The biosafety cabinet should not be used close to or stacked with other device. If it must
be used close to or stacked with other device, it should be observed and verified if it can
operate normally under the configuration.