21
4.3.2 Buttons
Password interface: after power on, the password interface will pop up. The initial password is 2222,
and the password can be modified. The modification method is shown below.
Button function: After pressing the corresponding function button, there will be icon display on the
display screen and the corresponding function will be started.
In case the fan button is pressed, there will be an animation demonstration after the fan runs; and the
fan will stop automatically when the glass door is closed.
Press the UV button to turn on the UV disinfection lamp and press it again to turn off the UV
disinfection lamp.
When the lighting button is pressed, the LED lamp is turned on, and when pressed again, the LED
lamp is turned off.
When the socket button is pressed, the socket is powered on, and if pressed again, the socket is
powered off.
Horn button: when the machine is in alarm state, press this button to cancel the alarm, and press it
again to start the alarm function.
Lock screen function: touch the lock screen button, the operation interface is in the locked state, at
this time other function buttons are in the locking state to prevent disoperation.
Interlock: UV lamp and fan are interlocked with each other. That’s to say, UV lamp cannot be turned
on when the lighting button is turned on; when UV lamp is turned on, the LED lamp can be turned
on and UV lamp is turned off; Note: when the fan is in normal operation, if the glass door is closed,
the fan will stop working, but there is an icon. After the glass door is lifted, the fan will resume
operation, and the UV function is the same as above.
Alarm window: when an alarm is triggered, the alarm information will be displayed in the
corresponding alarm window. When multiple alarms appear at the same time, it will be broadcast
circularly.
Button memory function: fan, LED lamp and lighting are provided with power-off memory function.
If the power is turned on after power failure, the corresponding function button will be opened,
where there is no need to re-input the password but enter the main interface.
User setting button: setting interface (click setting button in the main interface to enter the settings
interface).
Date and time adjustment: click the time in the upper right corner of this interface to enter the time
adjustment interface. To adjust the corresponding date and time in this interface, press the OK button
to save and exit.
Timing on / off function: click Time Mode button in the user settings interface to enter the setting
interface. Under this interface, the timing on/off function and timing shutdown function of fan, UV
lamp and socket can be set. For specific operation, input the time for timing on/off operation in the
corresponding function bar, and press the corresponding ON button after input to turn on the function.
At the same time, in the main interface, there will be an icon prompt in front of the corresponding
operation button, and the corresponding function will automatically be on/off according to the set
time. To turn off this function, click OFF button. Note: fan and UV lamp functions are interlocked by
glass doors.
View the working time of filter, UV lamp: click the Device Life button in the user settings interface
to enter the view interface to view the service life.
User password modification: in the settings interface, click the Change Password button to enter the
Содержание BSC-4FA2-HA