27
Cuvettes polluted
Clean the cuvettes
(6) Incorrect Reading
Possible Reason
Solution
Dark Current Error
Re-get the dark current
Worse matching of the cuvettes
Improve the matching of the cuvettes
3. Spare parts replacement
(1) Replace the Fuse
Danger! Be sure to switch off the power and unplug the socket before replacement!
Step 1, Tools Preparation
Prepare a 3×75 Flat Blade screwdriver
.
Step, Switch Off the Power Supply
Switch off the power supply, and unplug the socket.
Step 3, Take Out the Fuse Seat
Take out the fuse seat by the screwdriver(Fig.5-1).
Fig. 5-1
Step 4, Replace a New Fuse
Pick out the spare fuse and replace it to the working position(Fig.5-2).