16
4) Set Standard Sample.
According to the notice, input the standard sample concentration (Fig.
4-7),
to confirm, press
,
to select the value. After finishing inputting
concentration, press
to enter Abs. INPUT, the method is same as inputting concentration.
Then press
to enter curve picture, press
again to enter quantitative data sheet
to start test.
5) Calibrate 100%T/0Abs.
Put the Reference Sample in the light path and press
to
calibrate 100%T/0Abs.
6) Test Sample.
Put the test sample in the light path, press
to read value.
Method 3
:
Coefficient method (Input Regression Equation)
1) Quantitative Method.
Press
,
,
to select coefficient method (PARAM).
toconfirm.
2) Set Parameter.
According to the menu options, press numeric key to set up fitting method,
wavelength method, units and room, press
to confirm, then press
to enter,
input wavelength, photometry coefficient K1 and intercept K0 (Linear Fit)
to confirm.
3) Sample Test.
Press
to enter quantitative data sheet. Fig. 4-6
Fig. 4-6