Data Files
Open the data files saved on the SD card on the computer using Microsoft Excel or Notepad. The
figure below is an example data spreadsheet. Data values included are the date and time of the
measurement, the ethylene level in ppm, the mode (monitor, measure or cleaning, offset or GC
emulation), the CO
2
and O
2
concentration, the temperature of the gas stream in degrees Celsius, the
relative humidity (RH) of the gas stream in percent, the atmospheric pressure (KPa) and the flow
rate of the gas stream in ml/min. In this example, when the CO
2
or O
2
sensors are not being used,
the concentration values are zero. For the mode, monitor indicates measurements without
PolarCept, measure/cleaning indicate PolarCept, and offset indicates zero calibration.
Figure 13: Example data spreadsheet for a CI-900 measurement.
REMEMBER:
Always save the data files to the computer before making changes or starting analysis.