CI-900 Instruction Manual 8/10/2015
1554 NE 3
rd
Ave, Camas, WA 98607, USA Phone: (360) 833-8835
sales@
cid-inc
.com
www.cid-inc.com
25
11.
Press
Enter
to repeat and inject a sample or press Stop to exit to main menu. There is a
“cleaning” time between samples, which can be skipped by pressing Enter. During cleaning,
the ethylene sensor is consuming the ethylene from the sample. The instrument will wait
until the ethylene signal is less than 100 ppb before finishing the cleaning. Cleaning will
take longer with high concentration samples.
Note
: if injection port does not have a pressure-release valve on the OUT side, briefly remove
and replace the injection port from the CI-900 to manually release the internal pressure build-
up between injections.
Section 3: Sample Injection
12.
Follow the same steps, but this time, inject the sample gas using the same volume of sample
as for the calibration gas.
13.
Choose “No” by pressing Stop when asked “Set as calibration?” to view the ethylene
concentration on the display.
14.
Press the
Enter
to start another measurement. Press Stop to exit to the Main Menu.
15.
For ethylene readings, the procedure should be calibration injection, verification injection,
sample injection, then repeat, for most accurate measurements.
Results
Enter: Repeat Stop: Done
peak count 2729747
2.00ppm
CO2 0.03%
O2 21.1%
Uptime 00:52 Bat 90%
peak count 2729747
1.92ppm
CO2 0.05%
O2 20.4%
Uptime 00:58 Bat 85%