CI-900 Instruction Manual 8/10/2015
1554 NE 3
rd
Ave, Camas, WA 98607, USA Phone: (360) 833-8835
sales@
cid-inc
.com
www.cid-inc.com
52
View Voltage
The View>Voltage screen shows the current voltage of the instrument. Press the RIGHT arrow
when Voltage is highlighted to check the unit’s voltage. If any of these numbers are rapidly changing
after the instrument has warmed up, please contact technical support at
Press the LEFT arrow or STOP to exit back to the View Menu.
Voltage is set for each individual unit before leaving the CID Bio-Science factory and is board
specific. For default voltage values, please see the Production Test Check Sheet at the end of the
Instruction Manual for factory voltage settings. V_System is the voltage of the control board. The
voltage for V_Charger should increase when you connect the USB charger. V_3volt is the digital
voltage from the CPU and the V_Analog is the converted analog voltage.
View Battery
The View>Battery menu displays indicates on the top line if the unit’s Charging is On or Off, or if the
unit is connected via USB cable to a computer. The QBAT value shows the battery level in
hexadecimal follow immediately by the battery level in percent. VBAT shows the voltage of the
battery and TBAT indicates if the temperature of the battery is ok. The Charge shows the output
status of the battery charger. Press the LEFT arrow or STOP to exit to the View Menu.
When the instrument is connected to the charger, the top line will indicated Charge On. Charge Off
indicates that the unit is not charging. USB Host in the upper left corner indicates that the unit is
connected to the computer or that the power supplied by the charger is not full charging voltage
(unplug any other devices using the same charger).
The QBAT hexadecimal number should be changing and the battery percent increasing. Also, the
voltage will begin to increase slowly. The Charge will read “off” when no charger is plugged in and
“Complete” when finished charging. The Board refers to the hardware version of the unit. Features
Charge
Off
QBAT:
a413 60%
VBAT:
3.840V
TBAT:
OK
Charge:
Off
Board:
v6
View Voltages
V_System = 5.98v
V_Charger = 4.64v
V_3volt = 3.31v
V_Analog = 2.07v
Содержание CI-900