GC Emulation Mode Operation Procedure
Section 1: Warm up
1.
Turn on GC Emulation mode in the Setup>Measure menu. Press
Save
to save changes and
exit to the main menu.
2.
Enter the measure menu and wait for the instrument to stabilize.
3.
Follow the instructions on the display and connect the inline injection port. Press
Enter
.
4.
Enable or disable the Water Chamber to increase selectivity of C
2
H
4
measurement. Enable
the chamber only if CO
2
and O
2
measurements are NOT involved. Chamber In should be
filled at least half full with fresh distilled water. The distilled water in Chamber In should be
replaced with fresh DI water
every 10 injections
. Alternatively, monitor mode can be used
to clean the water.
SENSOR STABILITY
File
file01
Flow
200 mL
Initialization
Automatic
Environment Stabilizing
Please Wait...42%
Uptime 00:11 Bat 98%
Setup injection tube
then press Enter
Uptime 00:13 Bat 98%
Setup
Measure
Mode GC Emulation
Use Water Chamber?
(C2H4 only)
Yes: Enter, No: Stop