43
Setup Time and Date
To change the time and date on the CI-900, press the RIGHT key when “Time” is highlighted on the
Setup menu. This utility allows the user to set-up the instrument in different time zones or to
adjust the time after daylight savings.
To change the time on the instrument, use the UP/DOWN arrows to change the values. Use
the LEFT/RIGHT arrow to highlight the appropriate column. Then, make the appropriate
shift in time and press SAVE.
Setup Flow
The air flow of the CI-900 is an important factor for taking good measurements. The default flow
rate is 200 ml/min. Flow rate is adjustable from 100-500 ml/min.
A flow value below 100 ml/min can negatively impact sensor functionality. If measuring bottled
gas, a lower flow rate may be used in order to reduce gas waste. If measuring open air, a higher flow
rate (~300 ml/min) is acceptable.
The default flow rate is 200 ml/min
.
To make adjustments:
1.
Go to Setup>Flow and press the RIGHT arrow.
2.
Use the UP/DOWN arrows to adjust the value.
3.
The Set Flow value is found in ml/min to the far right.
4.
The Actual Flow value will change to match the set flow.
5.
Press SAVE to save changes and exit back to the Setup Menu.
6.
Press STOP to exit back to the Setup Menu without saving.
Setup Time
03 28 2013 16 : 33 : 24
Setup Flow
Set Flow 200
Actual Flow
200