S3 Cell Sorter Instruction Manual
| 35
4.3 Daily Startup
On a daily basis, the system requires approximately 20 min of warmup time before running
quality control or any samples. To warm up the system:
1. Move the loading stage into the wash position.
2. Click on the
Setup and Maintenance
tab.
3. Click
Startup
.
Automatically the Startup Settings window will pop up (Figure 32). Using the startup button, the
instrument can either be started up at the moment or a scheduled startup can be designated.
Startup Now is the default option.
4. Choose startup option in the Startup Settings window.
5. Click the checkmark box to proceed.
Note:
If scheduling an auto startup, the software will return to the main window and will not
start up the system until the scheduled time.
The Startup Status window will display the status of the startup process (Figure 33). The startup
process includes initializing the system, filling the fluidics lines, and warming up the lasers.
Fig. 31. Automatic startup message in login window.
Fig. 32. Startup Settings window.
Fig. 33. Startup Process windows.
Содержание 145-1001
Страница 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Страница 2: ......
Страница 8: ...S3 Cell Sorter Instruction Manual vi vi...
Страница 10: ...S3 Cell Sorter Instruction Manual viii...
Страница 22: ...12 S3 Cell Sorter Instruction Manual 12...
Страница 42: ...S3 Cell Sorter Instruction Manual 32...
Страница 60: ...50 S3 Cell Sorter Instruction Manual 50...
Страница 74: ...S3 Cell Sorter Instruction Manual 64...
Страница 82: ...72 S3 Cell Sorter Instruction Manual 72...
Страница 96: ...86 S3 Cell Sorter Instruction Manual 86...
Страница 112: ...102 S3 Cell Sorter Instruction Manual 102...
Страница 114: ...S3 Cell Sorter Instruction Manual 104...
Страница 116: ...106 S3 Cell Sorter Instruction Manual 106...
Страница 118: ......
Страница 119: ......