38 |
S3 Cell Sorter Instruction Manual
38 |
4.4.2 Side Streams
Following droplet setup, the side streams are adjusted to align with the sort tube positions and to
ensure they are not fanning. The streams may be viewed by clicking
View Streams
on the Home
tab. The View Streams window displays a charge phase and defanning slider bar (Figure 38).
Users can manually adjust the charge phase and defanning values by moving the slider bar. The
Enable Test Pattern button should be selected in order to apply a charge to the stream.
Users with administrator privileges may access
and modify these settings in the Administrator
tab. For more information, refer to Section 3.3,
Administrator Tab Toolbar.
CAUTION!
Optimal side stream
settings are calculated automatically
after performing a quality control
procedure. Adjusting any of these
values will change the optimal
settings and will affect the side
streams and sort.
Fig. 37. Droplet Monitor window.
Fig. 38. View Streams window with charge
phase and defanning slider bars highlighted.
Содержание 145-1001
Страница 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Страница 2: ......
Страница 8: ...S3 Cell Sorter Instruction Manual vi vi...
Страница 10: ...S3 Cell Sorter Instruction Manual viii...
Страница 22: ...12 S3 Cell Sorter Instruction Manual 12...
Страница 42: ...S3 Cell Sorter Instruction Manual 32...
Страница 60: ...50 S3 Cell Sorter Instruction Manual 50...
Страница 74: ...S3 Cell Sorter Instruction Manual 64...
Страница 82: ...72 S3 Cell Sorter Instruction Manual 72...
Страница 96: ...86 S3 Cell Sorter Instruction Manual 86...
Страница 112: ...102 S3 Cell Sorter Instruction Manual 102...
Страница 114: ...S3 Cell Sorter Instruction Manual 104...
Страница 116: ...106 S3 Cell Sorter Instruction Manual 106...
Страница 118: ......
Страница 119: ......