background image

4

LVP180914 Rev. D

www.bindicator.com

TECHNICAL SPECIFICATIONS: ROD CONFIGURATIONS

FUNCTIONAL

Power Requirements Universal

(± 10%), 120-240 VAC 50/60 Hz or 24-48 VDC

Power Consumption

4 W AC; 4 W DC

Fuse 

Slow Blow, 0.5 A 300 V (Not User Serviceable)

Operating Temperature

Electronics: Ordinary Location

-40° to 158° F (-40° to 70° C)

Electronics: Hazardous Location

-4° to 158° F (-20° to 70° C)

Electronics: Hazardous Location -T

-40° to 158° F (-40° to 70° C)

Process: Standard Temperature

-22° to 203° F (-30° to 95° C)

Process: High Temperature

-22° to 320° F (-30° to 160° C)

Outputs

Main Relay

8 A DPDT @ 240 VAC or 30 VDC (resistive)

Auxiliary Relay - ADVANCED ONLY

0.46 A SPDT @ 150 VAC or 1 A @ 30 VDC

PERFORMANCE

Pressure Rating

150 psi (10.5 kg/cm

2

) with 

3

/

4

” NPT; 50 psi (3.5 kg/cm

2

) with 1 

1

/

4

” NPT

Time Delay - STANDARD

Field Adjustable; Up to 6 seconds

Time Delay - ADVANCED

Field Adjustable; Up to 150 seconds

Fail Safe

Field Selectable; high/low level

Sensitivity - STANDARD

Minimum 6 lbs/ ft

3

 (96 kg/m

3

); Field Adjustable

Sensitivity - ADVANCED

Minimum 3 lbs/ft

3

 (48 kg/m

3

); Field Adjustable

Remote Distance

50 ft. (15 m) Max

PHYSICAL

Enclosure Material

Powder or Epoxy Coated Aluminum or 304 SS

Process Fitting

NPT 

3

/

4

”, 1

1

/

4

” ,1

1

/

2

” 

BSP R 

1

1

/

4

” ,1

1

/

2

” 

BSP G 

1

1

/

4

” ,1

1

/

2

” 

Rod Material

316 Stainless Steel

Dual Conduit Entry

3

/

4

” NPT or M20 x 1.5

Mounting Plate Material

Mild Steel, 304 Stainless Steel

Extended Pipe Material

316 Stainless Steel

Shipping Weight

Integral, non-extended 10 lb (4.5 kg)

Pollution Degree

2

Installation Category

II

Altitude

6,562 ft (2,000 m)

APPROVALS & RATINGS

UL:

Ordinary Location Type 4X; IP66 (US and Canada)

Hazardous Location, Type 4X

 

Explosion Proof, Class I, Div 1, Groups C, D (US Only)

 

Dust Ignition Proof, Class II, Div 1, Groups E, F, G (US and Canada)

CE

Electromagnetic Compatibility Directive

Low Voltage Directive

Содержание LP II Series

Страница 1: ...Pulse PointTM II Series Vibrating Rod Installation Operation Manual IOM LVP180914 Rev D...

Страница 2: ......

Страница 3: ...nfiguration Installation Operation Manual CONTENTS I HANDLING STORAGE 1 II GENERAL SAFETY 2 III PRODUCT DESCRIPTION 3 IV MECHANICAL INSTALLATION 5 V ELECTRICAL INSTALLATION 7 VI SET UP 14 VII MAINTENA...

Страница 4: ...ES CONDITIONS OR PROCEDURES WHICH IF NOT FOLLOWED COULD RESULT IN SERIOUS INJURY RISK OF ELECTRICAL SHOCK CAUTION IDENTIFIES CONDITIONS OR PROCEDURES WHICH IF NOT FOLLOWED COULD RESULT IN SERIOUS DAMA...

Страница 5: ...AND RECYCLING This product can be recycled by specialized companies and must not be disposed of in a municipal collection site If you do not have the means to dispose of properly please contact for re...

Страница 6: ...warranty reasons use of accessory equipment not recommended by the manufacturer or modification of this device is explicitly forbidden All servicing of this equipment must be performed by qualified se...

Страница 7: ...eater than the set threshold determined by the sensitivity setting the alarm relay will turn ON or OFF depending on the fail safe settings APPLICATIONS Unlike radio frequency and capacitance technolog...

Страница 8: ...e Up to 6 seconds Time Delay ADVANCED Field Adjustable Up to 150 seconds Fail Safe Field Selectable high low level Sensitivity STANDARD Minimum 6 lbs ft3 96 kg m3 Field Adjustable Sensitivity ADVANCED...

Страница 9: ...THE HOUSING OR ROD MAY DAMAGE THE UNIT AND VOID THE WARRANTY CAUTION WHETHER MOUNTING DIRECTLY THROUGH A SIDE WALL OR PIPE EXTENDED AND MOUNTED VERTICALLY THROUGH THE TOP OF A VESSEL NEVER ATTEMPT TO...

Страница 10: ...used with powders it should be installed vertically or at a downward angle that exceeds the angle of repose to reduce material build up see figure 1 Figure 1 Mounting Orientation Do not mount the rod...

Страница 11: ...l equipment you should always follow basic safety precautions including the following The installation and wiring of this product must comply with all national federal state municipal and local codes...

Страница 12: ...240 VAC 50 60 Hz or 24 48 VDC and provides reverse polarity protection in the event of a wiring error LP II SERIES INTEGRAL MODEL ONLY Input Power Connections 1 Refer to Figures 3 or 4 and 5 when conn...

Страница 13: ...n connecting input power to the unit 2 Loosen set screw that locks cover in place 3 Unscrew the housing cover and remove Note Two threaded female conduit openings are provided in the remote housing to...

Страница 14: ...h conduit and strip 3 16 4 to 5 mm 18 Connect factory supplied cable to terminal board as shown in Figure 10 19 Check that all wires are held tightly in place by lightly pulling each conductor 20 Repl...

Страница 15: ...THE GASKET CAN ALLOW MOISTURE TO ENTER THE ENCLOSURE AND DAMAGE THE UNIT Figure 4 LP II Integral Enclosure with Cover Removed ADVANCED Figure 5 Power and Ground Connections Figure 6 Main Relay Connec...

Страница 16: ...D www bindicator com Figure 8 LP II Remote Enclosure with Cover Removed STANDARD Figure 9 LP II Remote Enclosure with Cover Removed ADVANCED Note 1 For safety properly ground the enclosure to an adequ...

Страница 17: ...13 www bindicator com LVP180914 Rev D Figure 10 LP II Remote Connection Diagram...

Страница 18: ...will discuss configuration for optimum performance in a given application The settings are controlled by SW1 as shown in Figure 11 PRODUCT OVERVIEW ADVANCED MODEL Figures 4 and 9 show the electronics...

Страница 19: ...rm LED is ON Non Alarm State material below the rod Main Relay is energized Relay NC contacts are open Relay NO contacts are closed Alarm LED is OFF Note If the electrical power fails the main relay t...

Страница 20: ...or its absence and the output relay changes state is field programmable using SW5 Positions 2 and 3 The delay can be for when the product touches the rod or when the product leaves the rod or both reg...

Страница 21: ...s factory set for the lowest sensitivity The table below is for illustration purposes of a vertically mounted unit only and results will vary depending on material properties and conditions SW3 Sensit...

Страница 22: ...in the table below WARNING TO AVOID IGNITION HAZARD DUE TO ELECTROSTATIC DISCHARGE CLEAN ONLY WITH A DAMP CLOTH WARNING IN ORDER TO MAINTAIN SAFE OPERATION IN HAZARDOUS LOCATIONS THE INTEGRITY OF THE...

Страница 23: ...Ensure the cover is properly torqued and unit is installed wrench tight CODE CONDITION RELAYS 1 No or low excitation to the rod Aux Relay De energized Alarm Relay De energized 2 No return signal from...

Страница 24: ...150 Venture Boulevard Spartanburg SC 29306 Tel 800 778 9242 Fax 864 574 8063 sales bindicator com www bindicator com 2016 All rights reserved All data subject to change without notice LVP180914 Rev D...

Отзывы: