M (E2) 04/2019
page 9/71
DANGER
Electrical hazard.
Danger of death.
The chamber must NOT become wet during operation or maintenance.
The chambers were produced in accordance with VDE regulations and were routinely tested in accord-
ance to VDE 0411-1 (IEC 61010-1).
During and shortly after operation, the temperature of the inner surfaces almost equals the set-point.
CAUTION
The inner chamber, the door gasket, and the exhaust duct will become hot during
operation.
Danger of burning.
Do NOT touch the inner surfaces, the door gasket, the exhaust duct or the charging
material during operation.
1.6
Intended use
Drying and heating ovens with forced convection and advanced program functions M are suitable for
drying and heat treatment of solid or pulverized charging material, as well as bulk material, using the
supply of heat. They can be used for drying purposes but are specially designed for solving all the prob-
lems which occur during material and ageing tests.
The chambers are suitable for harmless materials. A mixture of any component of the charging material
with air must NOT be explosive. The operating temperature must lie below the flash point or below the
sublimation point of the charging material. Any component of the charging material must NOT be able to
release toxic gases.
Other applications are not approved.
The chambers are not classified as medical devices as defined by the Medical Device Directive
93/42/EEC.
Do NOT use the chambers for drying processes when large quantities of vapor would form and result in
condensation.
Due to the special demands of the Medical Device Directive (MDD), these chamber are not
qualified for sterilization of medical devices as defined by the directive 93/42/EWG.
Following the instructions in this operating manual and conducting regular maintenance work
(chap. 12) are part of the intended use.
The charging material shall not contain any corrosive ingredients that may damage the ma-
chine components made of stainless steel, aluminum, and copper. Such ingredients include
in particular acids and halides. Any corrosive damage caused by such ingredients is exclud-
ed from liability by BINDER GmbH.
Содержание M 115
Страница 58: ...M E2 04 2019 page 58 71 15 7 Dimensions M 53...
Страница 59: ...M E2 04 2019 page 59 71 15 8 Dimensions M 115...
Страница 60: ...M E2 04 2019 page 60 71 15 9 Dimensions M 240...
Страница 61: ...M E2 04 2019 page 61 71 15 10 Dimensions M 400...
Страница 62: ...M E2 04 2019 page 62 71 15 11 Dimensions M 720...
Страница 63: ...M E2 04 2019 page 63 71 16 EU Declaration of Conformity...
Страница 64: ...M E2 04 2019 page 64 71...
Страница 65: ...M E2 04 2019 page 65 71 17 Product registration...