M (E2) 04/2019
page 31/71
Program entry as set-point step (example)
W/°C
t/min.
01 02 03 04 05 06 07 08
0
50
100
150
200
250
30
120
180
380
Tolerance
minimum
Tolerance
maximum
On
Off
t/min.
Operation line 1 = air flap
Program table corresponding to the diagram above:
Program
section
Set-point
temp.
Fan
Section
time
Operation
line1
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01
50
100 %
00:30:00
Off
1
0
-1999
+9999
02
50
100 %
00:00:01
Off
1
0
-1999
+9999
03
150
100 %
01:30:00
On
1
0
-5
+5
04
150
100 %
00:00:01
On
1
0
-1999
+9999
05
250
100 %
01:00:00
On
1
0
-2
+2
06
250
100 %
00:00:01
On
1
0
-1999
+9999
07
100
100 %
03:20:00
Off
1
0
-1999
+9999
08
100
100 %
00:00:01
Off
1
0
-1999
+9999
Now enter the values of the above program table into one of the 25 program places of the controller MB1:
For rapid transition phases, do NOT program any tolerance limits in order to allow maximum
heating speed.
PGM
09:17:15 15.12.13
Содержание M 115
Страница 58: ...M E2 04 2019 page 58 71 15 7 Dimensions M 53...
Страница 59: ...M E2 04 2019 page 59 71 15 8 Dimensions M 115...
Страница 60: ...M E2 04 2019 page 60 71 15 9 Dimensions M 240...
Страница 61: ...M E2 04 2019 page 61 71 15 10 Dimensions M 400...
Страница 62: ...M E2 04 2019 page 62 71 15 11 Dimensions M 720...
Страница 63: ...M E2 04 2019 page 63 71 16 EU Declaration of Conformity...
Страница 64: ...M E2 04 2019 page 64 71...
Страница 65: ...M E2 04 2019 page 65 71 17 Product registration...