CAUTION
Danger of destruction of samples.
Carry out the emergency release only in emergency.
Ensure that a service technician is available and/or that the samples can be stored
cold.
Proceeding for emergency release
•
Turn off the freezer at its
main power switch and pull
the power plug, thereto pull
the locking button on the
plug (Figure 23).
•
Use a suitable screwdriver
to unscrew the Phillips
screw of the Emergency
release cover
•
Remove the emergency
release cover
•
Use a fork wrench SW8 to
unscrew both screw nuts
(M5).
•
A suitable fork wrench is
supplied. It is located at a
magnetic holder inside the
filter grille (D) behind the
upper edge.
•
Remove the lock bolt from
the door
•
Now you can open the unit
door.