background image

How to Use the Ergolift 2 

 

 

12

For a transfer into a seated position; use the 
handles on the back of the sling to position 
the patient when transferring into a chair.  
Hold the handles firmly as the sling lowers 
and the sling will tilt back to position the 
patient. As the boom lowers, guide the lift 
away from the patient (approx. 15 cm/6 in). 

Once the patient is properly seated and the 
straps are loosed, remove the sling from the 
lift. 

Slide the lift away from the patient. 

Sling may be removed from around the 
patient. 

Directional Lock 

The Ergolift 2 is equipped with a directional 
lock which allows a better manoeuvrability. 
To engage the feature, align the wheels and 
flip up the directional lock located over the 
caster, at the back of the right leg. To allow 
the lift to pivot, remove the directional lock. 

Emergency Stop Feature 

1.

 

The emergency stop feature stops the lift 
from functioning in case of an 
emergency. It does not affect the 
charging system, the lift will charge 
whether or not the emergency stop is 
engaged.  

2.

 

To activate the emergency stop, press 
down on the red button on top of the 
control box on the lift. 

3.

 

To deactivate the emergency stop, turn 
the button clockwise. 

 

 

Emergency Lowering Feature 

The Ergolift 2 has an emergency manual 
lowering feature in the event of an electrical 
or functional failure. 

If the lift malfunctions when a patient is 
being transferred, the emergency manual 
lowering device provides a safe way of 
lowering the patient.  

There must be a load applied on the actuator 
for the emergency lowering feature to 
function. 

To operate the emergency lowering feature: 

 

Find the red tab on the side of the motor on 
the lift. 

 

 

Move the patient over a bed or chair. 

Slowly pull upwards on the tab and hold in 
place until the patient is lowered onto chair 
or bed. 

Once the patient is lowered safely into a 
chair or bed, take the lift to a qualified 
technician to be serviced. 

Содержание Ergolift 2

Страница 1: ...Ergolift 2 Instructions for Use 001 06109 EN rev 0 February 2010...

Страница 2: ...e patented or patent pending Our policy is one of continuous development and we therefore reserve the right to make technical alterations without notice The content of this publication may not be copi...

Страница 3: ...Labels 9 COMPATIBLE SLINGS 10 HOW TO USE THE ERGOLIFT 2 11 Transferring a Patient 11 Directional Lock 12 Emergency Lowering Feature 12 Anti crush Feature 13 Charging the Battery 13 Integrated Charger...

Страница 4: ...r safety and must be read and understood to help prevent injuries BHM Medical strongly advises and warns that to avoid injuries that can be attributed to the use of inadequate parts only parts designa...

Страница 5: ...ts It is to be used for patients within the specified weight limit indicated on the lift and the specified environmental conditions described in this manual CAUTION It is not recommended to use the li...

Страница 6: ...age of the equipment concerned Consumables must be maintained in accordance with published Instructions for Use and the Preventive Maintenance Schedule Equipment Identification The unit s identificati...

Страница 7: ...e 93 42 EEC This symbol is accompanied by the product s number of the manufacturer This symbol is accompanied by the product s serial number of the manufacturer This symbol indicates separate collecti...

Страница 8: ...this equipment WARNING BHM Medical mobile floor lifts are specifically designed for BHM Medical slings and accessories Slings and accessories designed by any other manufacturer are prohibited and wil...

Страница 9: ...he equipment is used by trained staff Violent impact during transfers is avoided Non compatible slings are not used with the lift The sling is intended to be used for this lift and can take the weight...

Страница 10: ...edical attention For recycling and disposal of the batteries the rules according to the WEEE directive Waste Electrical and Electronic Equipment as well as local laws and regulations must be followed...

Страница 11: ...Slings For more information about loop slings and how to use them refer to the documentation that comes with the sling Hammock sling THA X THAI X Hammock 6 sling THA6 X THA6I X Quick Fit sling TIR X...

Страница 12: ...dants prefer to connect the leg straps first This applies in particular to patients with large thighs You may need to lower the spreader bar a little using the hand control Make sure that all straps a...

Страница 13: ...feature stops the lift from functioning in case of an emergency It does not affect the charging system the lift will charge whether or not the emergency stop is engaged 2 To activate the emergency st...

Страница 14: ...ber of charging cycles of the battery is largely dependent on the depth of discharge in each cycle The more the battery is drained the shorter its lifespan The life of the battery is also related to s...

Страница 15: ...aced around the patient according to the instructions above and the documentation that comes with the sling 2 Choose the color of the loop you will need for the position you will be placing the patien...

Страница 16: ...rd surface disinfectant wipes that are supplied impregnated with a 70 v v solution of Isopropyl Alcohol A rubbing action will be necessary when using the wipes to promote effective disinfection of the...

Страница 17: ...ss under the base 3 Unscrew the 7 bolts which hold the base cover use an 8 mm socket Remove the base cover 4 Take out the actuator s two clevis pins and remove it from the base 5 Disconnect batteries...

Страница 18: ...ower cord there is a drawer or magazine where there are fuses Pull the drawer out gently Pull the burnt fuse out of the drawer that is next to the charger see diagram below and discard it Take the rep...

Страница 19: ...clevis pins on each end of the damaged inoperable motor 5 Unplug the actuator and replace it by the new one 6 Reinstall the cover Lubrication Casters The lift is designed for minimum maintenance howev...

Страница 20: ...4 Line up the mounting holes in the new front caster assembly and the fork 5 Install the bolt through the fork and caster assembly and thighten securely with the locknut Annual Inspection The Ergolift...

Страница 21: ...be inspected by the user ITEM INITIALLY BEFORE EVERY USE EVERY 2 MONTHS EVERY 6 MONTHS EVERY YEAR LIFT DEVICE Has the battery been charged Plug the lift whenever it is not in use X Verify that the saf...

Страница 22: ...URN TO FACTORY X Cycle to ensure smooth quiet operation X X A complete inspection of the actuator must be done by a qualified technician X X CLEANING Clean the lifter X WARNING For individual home use...

Страница 23: ...record of servicing along with signature of authorized service agent X X X Check all electrical components including battery charges for correct operation X MAJOR SERVICE to be check by a qualified te...

Страница 24: ...mits a beep during utilization The unit may stop lifting but you can still lower it Batteries are low recharge them Charge indicator on the control box green red does not light up when the battery is...

Страница 25: ...equipment Radio transmitting device mobile telephones etc should never be used near the lift since they can interfere with the function of the lift Battery type Rechargeable sealed lead acid Battery...

Страница 26: ...Technical Specifications 25...

Страница 27: ...gned Record any repairs required Withdraw the patient lifter from service if inspection reveals that user safety is jeopardized in any way from use of the patient lifter TO BE COMPLETED AFTER EACH SER...

Страница 28: ...ion taken Date _________________ Inspected by ____________________ ____________________ Printed Signature Service Type Pre delivery Periodic inspection Minor Major Condition report Action taken Date _...

Страница 29: ...ion taken Date _________________ Inspected by ____________________ ____________________ Printed Signature Service Type Pre delivery Periodic inspection Minor Major Condition report Action taken Date _...

Страница 30: ...onents repaired or replaced are guaranteed for the remaining period of the initial warranty Limitations and Exclusions The warranty above does not apply to serial numbered products if the serial numbe...

Страница 31: ...delay in the performance of this warranty due to cause beyond the control of BHM Medical breach of contract negligence or strict liability exceed the cost of the product covered hereby BHM Medical ne...

Страница 32: ...001 06109 EN rev 0 February 2010...

Отзывы: