![BHM Medical Ergolift 2 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/bhm-medical/ergolift-2/ergolift-2_instructions-for-use-manual_2748440019.webp)
Disinfection, Cleaning and Maintenance
18
3.
Replace the actuator with a new one (see
replacement parts). Plug the actuator into
the control box and secure it in place
with the bolt and plate.
Base Actuator Replacement
1.
Press the emergency stop button to
ensure the lift will not accidentally
operate while you are working on it.
2.
Flip the lift onto the mast to access under
the base.
3.
Remove the base cover by unscrewing
the bolts with an 8 mm socket.
4.
Remove the two clevis pins on each end
of the damaged/inoperable motor.
5.
Unplug the actuator and replace it by the
new one.
6.
Reinstall the cover.
Lubrication & Casters
The lift is designed for minimum
maintenance; however, a six (6) months
check and lubrication should ensure
continued safety and reliability.
The casters MUST swivel and roll smoothly.
A food grade type grease (waterproof auto
lubricant) may be applied to the ball bearing
swivel of the casters once a year.
Refer to lubrication diagram for lubrication
points. Lubricate all pivot points. Wipe all
excess lubricate from lift surface.
1.
Carry Bar
2.
Boom Mounting Bracket
3.
Boom/Mast Mount
4.
Mast Mounting Bracket
5.
Leg Pivots