Resistance
Range
Resolution
Accuracy
600 Ω
0.1 Ω
± (1.0% + 5)
6 KΩ
0.001 KΩ
60 KΩ
0.01 KΩ
600 KΩ
0.1 KΩ
6 MΩ
0.001 MΩ
± (1.5% + 5)
60 MΩ
0.01 MΩ
± (3.0% + 10)
Max. Open Circuit Voltage
: < 0.7 V
Diode
Range
Description
Remark
The approx. forward voltage drop of the diode
will be displayed.
Open Circuit Voltage: about 3V
Test Current: about 0.8mA
Continuity
Range
Description
The built-
in buzzer will sound if the resistance is less than about 20 Ω
The buzzer mayor may not sound if the resistance is between 20 and 150
Ω.
The buzzer will not sound if the resistance is more than 150
Ω
Open circuit voltage:
about 0.7V
Capacitance (use Relative Mode)
Range
Resolution
Accuracy
40 nF
0.01 nF
± (3.5% + 20)
400 nF
0.1 nF
± (2.5% + 5)
4 µF
0.001 µF
± (3.5% + 5)
40 µF
0.01 µF
± (4.0% + 5)
400 µF
0.1 µF
± (5.0% + 5)
4000 µF
1 µF
not specified
Frequency
Range
Resolution
Accuracy
9.999 Hz
0.001 Hz
± (1% + 5)
99.99 Hz
0.01 Hz
999.9 Hz
0.1 Hz
9.999 KHz
1 Hz
99.99 KHz
10 Hz
999.9 KHz
100 Hz
9.999 MHz
1 KHz
not specified
Input Voltage
: 1V rms - 20V rms
Temperature
Range
Resolution
Accuracy
°C
- 20°C – 0°C
0.1°C
± (6.0% + 5°C)
0°C – 400°C
0.1°C
± (1.5% + 4°C)
400°C – 1000°C
1°C
± (1.8% + 5°C)
°F
-4°F – 32°F
0.1°F
± (6.0% + 9°F)
32°F – 752°F
0.1°F
± (1.5% + 7.2°F)
752°F – 1832°F
1°F
± (1.8% + 9°F)
Note:
1. Use K type thermocouple.
2. Accuracy does not include error of the thermocouple probe.
3. Accuracy specification assumes ambient temperature is stable to ±1°C
For ambient temperature changes of ±5°C, rated accuracy applies after 1 hour