6
• Keep the charging interface away from something conductive such as copper
scrap.
• Please use original charger to charge the battery.
• Connect the handpiece to the charger base, the handpiece LED will indicate the
charging state, if not, please refer to the trouble shooting.
• There’s electronic circuit which control the torque (TORQUE LIMITER Function)
to prevent files from breaking; however, files may still break due to metal fatigue
if the torque is conditioned to be higher. Please check the working instruction of
file before use
CAUTION
• Fully training to use this product to guarantee the safety of patient.
• The product is used for dental treatment only by qualified personnel.
• Do not use a bent, damaged, deformed or non- ISO-conforming file. It may cause
personal injury due to sudden breakage or flying off during rotation.
• Does not use or leave the product in a high-temperature environment such as
under strong direct sunlight.
• Check the product before use, if there’s abnormal loose, vibration, noise and
temperature (heat generation). Immediately stop using and contact with your
dealer.
• Keep the handpiece away from dirt, it will affect the concentricity of contra angle
and the judgment of the torque.
• Before changing the contra angle or file or Maintenance of motor, turn off the
power of the motor.
• Do not lubricate any part of the motor except the contra angle.
• If use corrosive solutions during the operation, please clean the motor handpiece
immediately after operation.
• This equipment is for indoor use only.
• Follow the instruction of file manufacture to set the rotation and speed of the
motor.
NOTICE
• The motor use the rechargeable li-ion battery.
• The power consumption is low when the power is off, please fully charged after
each use, please charge at least 1 time a month if long-term use.
• When the battery is low, the motor automatically shut down, leave it for a while
and the motor can be turned on again, but this is not permitted, it may damage
and reduce the battery life.
• If the product is not used for a long period, please take the battery out.
• If the power is not left completely, the battery may be damaged, please replace a
new one.
• No need for the training of this product, please read the instruction before
operating and maintenance.