30
guidance and manufacturer’s declaration — electromagnetic emissions
The EndoPilot is intended for use in the electromagnetic customer or the user of the EndoPilot
shouId assure that environment specified below. The is used in such an environment.
Immunity test IEC60601 test level Compliance Level Electromagnetic environment - guide
Conducted RF
IEC 61000-4-6
Radiated 3RF
IEC 61000-4-
3 Vrms
150 kHz to
3V/m
80 MHz to
3 Vrms
3V/m
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the EndoPilot,
including cables, than the recommend-
ed separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
80 MHz 800 MHz
800 MHz~2.5 G Hz
Where P is the maxi mum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommend-
ed separation distance in meters (m).
Field strengths from fixed RF transmit-
ters as determined by an electromag-
netic site survey, shouId be less than
the compliance level in each frequency
range. Interference may occur in the
vicinity of equipment marked with the
following symbol:
NOTE 1 : At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 : These guidelines may not apply in at I situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (celIular/ cord less)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broad-
cast can not be predicted theoretically with accuracy. To assess the electromagnetic environ-
ment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the EndoPilot is used exceeds the applicable
RF compliance level above, the EndoPilot should be observed to verify normal operation. If ab-
normal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the EndoPilot .
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.