20
6.2 Initialization of Program (Factory-set configuration)
The program can be returned to the original state configured at the time of factory
shipment, if setup becomes confusing.
1. Turn off the power, when the power is on.
2. Press PROgRAM, and then press the POWER key for more than one second.
Release the keys when the LCD panel indicates "RES”. The machine on hold.
3. Re-press the PROgRAM key more than 1 second, lease the key when display “FIN”.
4. after the long beep, reset success.
Caution
If this function is used, all programs will disappear and return to the originally set
values. Record the present program details if required before you carry out this
operation.
6.3 Calibration
This function is to decrease fluctuation in the rotation speed of the motor handpiece and
the difference in torque by the contra angle.
1. Lubricate the contra angle head. (Refer to the “7 (1) Lubricating Contra Angle.")
2. Turn off the power. (If the motor power is on)
3. Press PROgRAM, and press the POWER key for more than one second. Release
the keys when the LCD panel indicates “RES ” And then press “+” key three times.
4. The LCD panel displays “CAL" with the alarm sound. Attach the lubricated contra
angle to the motor.
5. Put the handpiece into the charger, and then press ON/OFF key.
6. If the motor handpiece begins to rotate, leave it as it is until it stops.
7. This process ends, if the rotation stops and the LCD panel display returns to its
original state.
8. If you want to stop this process, turn off the power key for more than 1 second.