4
2. Warning and safety
Warning
•
First read this instruction manual before operating this device
•
This device should only be operated by licensed dental professionals
•
Do not directly or indirectly place this device near heat source.
•
Operate and store this device in reliable environment.
•
This device requires special precautions regarding electromagnetic compatibility
(EMC) and must be in strict accordance with the EMC information for installation
and use. Do not use this equipment especially in the vicinity of fluorescent lamps,
radio transmitting devices, remote control devices, handheld and mobile high-
frequency communication devices.
•
Extended time use of Reciprocating Mode may result in motor overheat
•
Use the original contra angle
•
Do not make any changes to the device. Any changes may violate safety
regulations, causing harm to the patient
•
Use original power adapter. Other power adapter will result in damage to lithium
battery and control circuit.
•
The motor handpiece cannot be autoclaved. Use disinfectant of neutral pH value
or ethyl alcohol to wipe its surface.
•
Do not remove contra angle before motor stops. It would cause damage to
handpiece and contra angle.
• Secure the file before starting
• Set torque and speed as per the recommended specifications of file
manufacturer.
•
Follow instruction in this manual to replace battery and only replace with original
lithium battery.
•
Remove battery from the handpiece before long-term storage
•
Do not touch handpiece or charging base for more than 10 sec during charging.
Heat builds up while charging.
Safety
• Do not use device in the presence of a flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
•
Do not continuously use contra angle with patient longer than 10 minutes. The
temperature of contra angle may reach 46.6°C. Allow time to cool down before
further use.