29
Table 3: Guidance & Declaration - electromagnetic immunity concerning
Conducted RF & Radiated RF
Guidance & Declaration - Electromagnetic immunity
The model ApexPilot G2 is intended for use in.the electromagnetic environment
specified below. The customer or the user of the models
ApexPilot G2 should
assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted
RF lEC
61000-4-6
Conducted
RF lEC
61000-4-6
Radiated
RF lEC
61000-4-3
3 Vrms
150 kHz to
80 MHz
6 Vrms ISM
frequency
band
3 V/m
80 MHz to
2.7 GHz
3V
6V
3V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of
the models ApexPilot G2, including
cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.2×P1/2
d=2×P1/2
d=1.2×P1/2 80 MHz to 800 MHz
d=2.3×P1/2 800 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter
In watts (W) according to the
transmitter manufacturer and d Is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,a
should be less than the compliance
level in each frequency range.b
Interference may occur In the vicinity
of equipment marked with
the following symbol:
NOTE I
At 80 MHz end 800 MHz. the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.