7.Troubleshooting
Failure
Possible cause
Solutions
Continuous beeping after
starting the handpiece
The continuous beeping
indicates that the hand-
piece is under reverse
rotation state.
Stop the handpiece and
change the operating
mode to Continuous
Rotation Mode.
Contra-angle calibration
failure
Calibration failure caused
by strong resistance of
contra-angle
Calibration failure caused
by strong resistance of
contra-angle
Clean the contra-angle,
and recalibrate after oil
injection.
Handpiece heating
Extended usage in
Reciprocating Motion
Mode
a) Normal phenomenon.
b) Stop use. Use after
the temperature of
handpiece drops.
After plugging the
handpiece into charging
base, the wireless
charging indicator light
does flash
a) The handpiece is not
seated in place.
b) The handpiece is fully
charged.
Reseat the handpiece in
place
The time of endurance
becomes shorter after
charging
Battery capacity is
reduced
Please contact
local distributor or
manufacturer.
The continuously rotating
file is stuck at the root
canal
Incorrect specification
setting.
Too high load torque of
file.
Choose Reverse
Rotation Mode, start the
handpiece, and take the
file out.