9
Standby
Memory location M3
Forward mode – clockwise arrow
Battery level – not fully charged
Speed: 300rpm
Torque: 3.0 Ncm
Operation
Torque scale
Torque setting at 3.0 Ncm
Real time torque shows below 1.0 Ncm
4.4 Contra-angle calibration
Before using right after purchase, whenever the motor handpiece or contra angle
has been replaced, or if the motor alternates between forward and reverse rotation
outside the canal, calibrate the unit in the following way:
Press “S” and “-“ and hold for 2 seconds to start calibration
After calibration is complete, the display screen will show “succeed” and will then
return back to Standby Display (initial display).
Note:
Plug in original contra angle without file.
Warning:
•
Use original contra angle only.
•
Any load applied to contra angle will be offset and calibration will not be correct.
• Use caution when inserting and removing files to avoid injury to fingers
5. Screen Display
5.1 Standby and operation