8. Cleaning, Disinfection and Sterilization
8.1 Foreword
For hygiene and sanitary safety purposes, the contra angle must be cleaned,
disinfected and sterilized before each usage to prevent any contamination. This
concerns the first use, as well as all subsequent uses.
8.2 General recommendations
8.2.1 Use only a disinfecting solution which is approved for its efficacy (VAH/
DGHM-listing, CE marking, FDA and Health Canada approval) and in accordance
with the DFU of the disinfecting solution manufacturer.
8.2.2 Do not place the contra angle in a disinfectant solution or in an ultrasonic
bath.
8.2.3 Do not use chloride detergent materials.
8.2.4 Do not use bleach or chloride disinfectant materials.
8.2.5 For your own safety, please wear personal protective equipment (gloves,
glasses, mask).
8.2.6 The user is responsible for the sterility of the product for the first cycle and
each further usage as well as for the usage of damaged or dirty instruments where
applicable after sterility.
8.2.7 The water quality has to be convenient to the local regulations especially for
the last rinsing step or with a washer-disinfector.
8.2.8 Do not sterilize the motor handpiece, the AC adapter or the base. After each
use, all the objects that were in contact with infectious agents should be cleaned
using towels impregnated with a disinfecting and detergent solution (a bactericidal,
fungicidal and aldehyde free solution) approved by VAH/DGHM-listing, CE marking,
FDA and Health Canada.
8.2.9 To sterilize the endodontic files, refer to the manufacturer’s instructions for
use.
8.2.10 The contra angle needs to be lubricated after cleaning and disinfection, but
before sterilization.
8.3 Step-by-Step Procedure
#
Operation
Operating Mode
Warning
1
Preparation
Remove the
contra- angle from
handpiece
and charging base.