Beurer GL44
19
Evaluating critical measured values
Display
Blood glucose
Action
Very low blood glucose
level
under 20 mg/dL (under
1.1 mmol/L)
Seek medical attention
immediately.
Low blood glucose level
under 70 mg/dL (under
3.9 mmol/L)
Have a suitable snack.
Follow your GP's instructions.
High blood glucose level
on an empty stom-
ach, above 100 mg/dL
(5.6 mmol/L)
2 hours after a meal
over 140 mg/dL
(7.8 mmol/L)
If this high value persists 2 hours
after your last meal, this may in-
dicate hyperglycaemia (high blood
glucose). Seek medical attention
to coordinate any measures, if ap-
plicable.
High blood glucose level,
possibly ketones
over 240 mg/dL
(13.3 mmol/L)
Perform a ketone test. For this pur-
pose, seek medical attention.
Very high blood glucose
level
over 630 mg/dL
(35.0 mmol/L)
Take another measurement using a
new test strip. If the display is the
same as before: seek medical as-
sistance immediately.
5.6 Functional check using control solution
The control solution is used to test the entire blood glucose monitoring system. This helps to deter-
mine whether the monitor and the test strips are working optimally together and whether the test is
being performed correctly.
Perform the control solution test if you suspect that the monitor and/or the test strips could be faulty
or if you have repeatedly measured unusual blood glucose values. Also test the monitor if it has been
dropped or is damaged. The control solution is available separately. For the control solution test,
please observe the additional notes in the instructions for use for the control solution.
Important
• Do not use control solutions by a different manufacturer. Correct functioning of your monitor can
only be tested using Beurer control solutions ( LEVEL4).
Содержание GL 44