28
Beurer GL44
8 w
hat
if
there
are
problems
?
Display messages on batteries and blood glucose measurement
No.
Cause
Solution
LP
Batteries empty.
Replace all batteries.
Ht
Temperature of the measuring envi-
ronment, monitor or test strip above
the permitted range.
Repeat the test using a new test strip as soon as
the measuring environment, monitor and test strip
have reached room temperature (+20°C to +26°C).
The check performed at room temperature is used
as a general functional check. The operating range
specified in the technical specifications is valid
without restriction.
Lt
Temperature of the measuring envi-
ronment, monitor or test strip below
the permitted range.
Repeat the test using a new test strip as soon as
the measuring environment, monitor and test strip
have reached room temperature (+20°C to +26°C).
The check performed at room temperature is used
as a general functional check. The operating range
specified in the technical specifications is valid
without restriction.
Err
Used or contaminated test strip in-
serted.
• Insert an unused test strip that has not expired.
• Repeat the blood glucose measurement.
Err001 System error.
Remove batteries, reinsert batteries. Should the
problem persist, contact Customer Services.
Err002 Insufficient blood on the test strip.
Repeat the measurement using a new test strip.
Err003 The haematocrit value is outside the
range 20–60%.
Repeat the measurement using a new test strip.
Should the problem persist, contact Customer
Services.
Err005 System error.
Remove batteries, reinsert batteries. Should the
problem persist, contact Customer Services.
Unknown error messages.
Remove batteries, reinsert batteries. Should the
problem persist, contact Customer Services.