BESANTEK
- - - - - -
•
Ices OFF
•
Re
Main Status Indicators by LED:
Constant Current Source Indicator.
When this LED lights up, the Constant Current
Source is shut down.
The Constant Current Source is shut down when
the test is not in progress.
While in progress, the Constant current Source
LED does not lights up.
After Switching the Instrument ON, the Ices OFF
led will lights up.
When a test has stopped, the Ices OFF led will
lights up.
In case of Over-Temperature, the Ices OFF led will
lights up.
The Re indicator LED lights up, when the
resistance between the current terminals is too
high for the constant current source to regulate.
This indicator is only significant while the test is in
progress. You can try to lower the test curent if it
comes on while testing.
The RP indicator LED lights up, when the
potential between the potential terminals is too
high. This also can be due to the resistance under
test being too high for that current range. Try
to lower the current or increase the resistance
range.(only significant while the test is progress).
-9-