-4-
The I
CCS
LED is the LED indicating the Constant Current Source
being OFF. When this led lights up, it mean that the current injection
has been switched OFF (could be over-temperature).
The METER has a built-in custom 3½ digit liquid crystal display
which can be viewed in most lightning conditions.
This display indicates the conditions (Hold, m, buzzer, polarity
condition of load, + or -, automatic decimal point change with range
selection).
The ranges are selected by the 'SCROLL / RANGE' push button.
A test is initiated by pressing the TEST/STOP push-button.
The instrument takes measurements for 10 seconds if the TEST/
STOP push-button is depressed for less than 3 seconds (EnerSave
mode).
If the TEST/STOP push button is depressed for more than 3
seconds, the test will carry on for 60 seconds.
The tester switch "OFF" completely when the OFF push button is
pressed, but The METER has also an auto-off feature
The tester "Hold" the last reading before stopping the test.
The tester is fuse protected and has a crowbar between C
1
and C
2
.
This crowbar is activated by voltage. If the voltage is too high, that
crowbar will blow the C
F
fuse automatically to interrupt the circuit.
The voltage between P
1
and P
2
is also protected for over voltage
.If that voltage is too high, the crowbar will blow the P
F
fuse
automatically to interrupt the circuit.
The tester has a temperature shut down.
The temperature sensing is done on the current regulation transistor.
Should this over-temperature alarm be activated, the I
CCS
OFF LED
will lit, indicating the current has been shut down.
Allow the instrument to cool down for a while before proceeding
further.