Cleaning and Maintenance
LB 983 NightSHADE II Operating Manual
112
8.2.4 Replacing the halogen lamp
The halogen lamp is located close to the excitation filters. It is protected by a black cover plate.
To replace the lamp proceed as follows. Make sure that the same type of lamp is used to guar-
antee proper light output.
Be careful when touching the lamp since it could be hot when the lamp was
in use before.
Switch off the instrument and unplug the mains connector. If the lamp was in use before
allow some time to cool down the lamp.
Open the top flap to get access to the lamp module.
Pull out the black protection cover to get access to the lamp.
Remove the lamp from the snap in holder.
Unplug
the
lamp
socket
and
plug
in
the
replacement
lamp.
Do not touch the reflector of the lamp to avoid damage of the lamp.
Place the lamp back into the snap in holder. Make sure the lamp is fitted properly in the
holder (round edge fit with lamp reflector).
Insert the protection cover. Make sure the cover is properly fixed and both pins are se-
curely fastened in the mounting holes.
Close the top flap.