7 Maintenance and Repair
Shields for Flange Mounting
46
44192BA2, Rev.03, 07/2021
2.
Provide the transport shield with the source to be installed.
3.
Open the transport shield.
4.
Use the pliers to pull the source out of the transport shield.
1
2
3
Mounting flange
Source
Rod for pushing the point source or rod source
Fig. 22
Installing the source in the shield (left: point source, right: rod source)
5.
Insert the source into the shield as quickly as possible with the coupling part
of the source holder.
6.
Carefully push the source with a rod further in the direction of the shield
head until it stops.
7.
Secure the source with the corresponding clamping screw.
8.
Close the shutter.
9.
Mount the shield again on the dip pipe, see chapter 5.3 Installation of the
Shield.
10.
Bring the source into the measuring position, see chapter 6.2 Commissioning.
The removal and installation of the source is completed.