QUICK CHECK (Cont'd)
If you have a 19” chassis (BPM-RFC/X), use the following set-up:
DVM
DVM
Y
X
RF Source
Power
splitter
Oscilloscope
Ch.1 Ext. trig.
To button inputs
(back of chassis)
Attach the equipment together as shown above.
Set the oscilloscope time base on 0.2 ms / div.
Ext. trigger connects to SYNC signal, trigger level 0.2V, 1 M
Ω
AC coupling
Channel 1 to SDEMOD signal, sensitivity 0.1 V / div., 1 M
Ω
AC coupling
Set the RF source to the operating frequency.
Amplitude ca. -13 dBm
Note: The BPM module operating frequency is written on the F–Key daughter board.
To access the F–Key, remove the BPM module shield.
Set the X and Y voltmeters range in such a way that millivolts are readable.
Connect to AC mains, the oscilloscope should display:
Synchronous demodulated signal.
Each period 100…125µs represents
the successive buttons A, B, C, D, A, B…
The signals are equalized before delivery
of the BPM module, using a precision
input 4-way splitter.
Differences in signal level as shown here
are due to uneven input signals.
Note: 1 dB input signal difference gives
ca. 300 mV of demodulated signal
amplitude difference
BERGOZ Instrumentation
Multiplexed
01630 Saint Genis Pouilly, France
Beam Position Monitor
Tel. +33-450.426.642
Fax +33-450.426.643
Page 4
User’s manual
BERGOZ Instrumentation - 01630 Saint Genis Pouilly, France - Tel.: +33-450.426.642 - Fax: +33-450.426.643
TVA-VAT-IVA-USt. Nº FR88414997130 - Sàrl. capital 152K - Siren 414 997 130 R.C.S. Bourg - APE 332B