Device description
7KHVSHFL¿HGGLVSOD\DQGRSHUDWLQJHOHPHQWVDUHGHVLJQDWHGDVIROORZV
1
Irradiation sensor
2
Digital display
3
-key,
temperature
4
-key,
angle of inclination
5
-key,
time/date
6
Battery compartment
(rear)
7
-Tkey,
HOLD/ OK/ LOG
8
HOLD indicator
9
Marking for compass bearing
J
USB jack
K
Jack (Link),
for other BENNING products
L
Jack (PROBE),
temperature sensors
M
Ambient temperature sensor
N
Module temperature sensor
Measuring the insolation, the angle of inclination and the compass
bearing
Insolation
Ź
Angle of inclination
Ź
Compass bearing
Ź
Insolation
1. Switch the device ON and place it onto the surface of the photovol-
taic module.
2. The insolation is displayed in the unit W/m² or BTU/hr/ft².
Angle of inclination
1. Press the
4
key for angle of inclination / compass bearing.
2. The angle of inclination with respect to the horizontal is indicated
on the LC display.
Null balance of the angle of inclination
1. Place the device onto a plane surface.
2. Press the
4
key for > 5 seconds until the "Z" symbol starts
flashing on the LC display.
3. Press the
4
key again to store the "+SET" value.
4. Turn the device by 180° and press the
4
key again to store the
"-SET" value. The angle of inclination has been balanced to null.
The "Z" symbol is shown on the LC display.
5. Press the
4
key again for > 5 seconds to exit the setting. The
"Z" symbol disappears from the LC display.
Compass bearing
1. Press the
4
key for angle of inclination / compass bearing.
2. To determine the cardinal direction, hold the device horizontally. As
soon as the angle of inclination e/- 20°, "- - -" appears on
the LC display and it is not possible to take a bearing.
3. Hold the marking
9
on the top of the device in the direction which
you would like to determine. The direction is displayed in degrees:
0° = North, 90° = East, 180° = South, 270° = West
4. Please observe that metallic objects or live conductors might affect
the measuring accuracy of the device.
Measuring the insolation, the module temperature and the ambi-
ent temperature
Insolation
Ź
PV module temperature
Ź
ambient temperature
Ź
Insolation
1. Switch the device ON and place it onto the surface of the photo-
voltaic module.
2. The insolation is displayed in the unit W/m² or BTU/hr/ft².
Module temperature and ambient temperature
1. Connect the temperature sensors
and
to the
BENNING SUN 2
.
2. Press the
3
key for temperature measurement.
3. Bring the module temperature sensor
into contact with the PV
module and wait until the temperature sensor has measured the
temperature of the PV module.
Changing the displayed unit
1. Press the
3
key for > 5 seconds until the displayed tempera-
ture unit starts flashing.
3UHVVWKHŹ
3
key to select the respective unit for insolation or
temperature.
3UHVVWKHŸ
4
RUź
5
key to change between the units °C or
°F or between the units W/m² or BTU/hr/ft².
4. Press the
7
key (OK) to store the setting.
SUN 2
ON/OFF
HOLD
OK
LOG
ON/OFF
HOLD
OK
LOG
02/ 2012
BENNING SUN 2
3
02/ 2012
BENNING SUN 2
2
02/ 2012
BENNING SUN 2
4