7. Preventive Maintenance/System Operational Check-Out
7-84
For your convenience, two preventive maintenance schedules have been provided.
Schedule 1: should be performed by either the Clinical User or a Biomedical Technician
(BMET).
Schedule 2: should be performed by either a BMET or other qualified service personnel.
Schedule 1
To be performed by either the Clinical User or a Biomedical Technician (BMET).
Routine Maintenance
Interval
Before or
After Each
Use
Every
Month
Every 6
Months
1. Clean and/or Disinfect Exterior, if necessary.
•
2. Clean Fluid Out and In-Line Air Detector.
•
3. Check the Power Cord.
•
4. Clean Temperature Probes
•
5. Check/Clean the Fan Guard.
•
6. Check the System Seal.
•
7. Check Instrument Door and Ceramic Disk.
•
8. Check Rubber Feet.
•
A.
SERVICE AND
PREVENTIVE MAINTENANCE SCHEDULE