6. Setting System Parameters and Systems Calibration
6-72
Listed below are System Parameters Setup and Hardware Calibration. These
calibrations, pressure transducer, temperature probes, and power module, need to be
calibrated only after verifying that it is operating outside accepted parameters. See
Hardware Verification, Chapter 7.
WARNING
!
Do not access system parameters setup while the instrument is
patient connected.
Changes in system setup can be made to:
1.
Date and time
: Set the real time clock and date
2.
Display brightness
: Change the display brightness
3.
Key Rate
: Set touch key sensitivity
4.
Bolus volume
: Set the bolus delivery volume
5.
Pressure limits
for High Pressure alarm: Set the maximum allowable in-line
pressure. The possible setting range from 100 - 300 mmHg.
SYSTEM PARAMETERS SETUP