6. Setting System Parameters and Systems Calibration
6-78
Pressure Calibration
:
1.4.1 Press PRESS CAL. Enter the password 013192.
1.4.2 Close the bag clamps and block the air vent on top of the reservoir
chamber.
1.4.3 Disconnect the patient line and connect the pressure source to the
luer fitting at the patient line port of the disposable set and apply
pressure while monitoring the amount of pressure with a
manometer.
1.4.4 Apply 100 mmHg of pressure. Press UPDATE to accept the 100
mmHg calibration point.
1.4.5 Apply 258 mmHg
±
of pressure. Press UPDATE to accept the
calibration point. The system will return to the main
Calibration/Setup screen.
1.5
Calibration Verification:
1.5.1 Press HARDWARE. Enter the password 013192.
1.5.2 Press RIGHT VALVE to set the valve in the right position.
1.5.3 Apply 258 mmHg into the disposable. The pressure status line
should read 258 mmHg (± 25 mmHg).
1.5.4 Repeat with 100, and 300 mmHg pressure. Tolerance limit is ± 25
mmHg
1.5.5 System is now pressure calibrated and verified.