Chapter 4: Parameters Setting, Service and Preventive Maintenance
4-7
7.
Seals
Inspect the seal around the unit to make certain it is in good condition. Check also
the seal around the touch screen and ceramic disks. Use Dow Corning 732
multipurpose RTV sealant or equivalent if needed to maintain fluid resistance.
8.
Instrument Door and Ceramic Disks
The instrument door must fit properly for the system to operate correctly. The
platen part of the roller pump is located on the door. The platen must line up
properly with the pump.
a.
Check hinges for debris build-up, clean any dried debris/fluid from hinges
area. Be sure that door is seated completely down on the hinges.
b.
Check plastic rivets and door integrity. Make sure that the doorframe is not
bent. Replace, if bent.
c.
Inspect the ceramic disks on the door and in the center of the unit for
cracks. Return to manufacturer for replacement if they are damaged.
9.
Rubber Feet
Inspect the rubber feet on the bottom of the unit for cracked or missing rubber feet.
Replace if necessary.