Chapter 4: Parameters Setting and Preventive Maintenance
4-4
2.
Display Brightness
There are four (4) levels of display brightness. Press DISPLAY BRIGHT to change
the present level of brightness to the next level.
3.
Language Setup
Press this key to set screens to your preferred language.
4.
Key Rate
The key rate sets up the sensitivity of the touch keys. There are three different
levels of sensitivity; FAST, MEDIUM and SLOW. The current level of sensitivity is
indicated on the key itself. The FAST setting requires the least amount of time for
a key to respond. The MEDIUM setting requires more time and the SLOW key
requires the most time and makes the touch keys least sensitive.
The key
sensitivity is set at factory to Medium
.
Note that this key changes the time required to depress a key for stroke to be
recognized. The pressure required is not affected.
5.
Pressure Limit
The user can set the maximum allowable in-line pressure. The possible setting
ranges from 100 to 300 mmHg. The current pressure limit value is displayed on
the PRESS LIMIT status line on the Calibration/Set-Up screen. Press and hold
the key to change the limit in increments of 50 mmHg. During infusion, the system
keeps the pressure in the line under the pressure limit by reducing the infusion rate
as the in-line pressure approaches the pressure limit.
The pressure limit is
automatically reset to 300 mmHg each time that the system is powered on.