Chapter 2: Operation
2-6
INSTALL 4.4 LITER RESERVOIR AND THE
STRAIGHT INFLOW/ BIFURCATED OUTFLOW
PATIENT LINE
Straight Inflow/Outflow Patient Line Configuration
with 4.4 Liter Reservoir, Heat Exchanger and
Vacuum Regulator
Reservoir Top View
4.4 Liter Reservoir with Straight Inflow/
Bifurcated Outflow Patient Line
Version
1.
Place the reservoir into the holder.
2.
Assemble the reservoir, using aseptic
techniques, as follows:
Top of Reservoir
: Remove all vented caps from
top of reservoir and install these parts to the
marked locations:
Bag Spike for STERILE SOLUTION
Tubing with injection port to LUER
Regulator source tubing from the vacuum
trap to VACUUM
3. Using aseptic technique, connect the Reservoir
outlet, 3/8” O.D tubing, and the Recirculation
Line, 3/16” O.D tubing, to the connectors of the
Heat Exchanger.
4.
Adjust the Reservoir Holder to make sure that
the two connection leads underneath the
reservoir are not stretched or kinked.
Stretched or kinked connection leads can cause
flow restrictions.
5. Close all clamps and the ON/OFF pinch clamps.
INSTALLING PATIENT RETURN LINE
1.
Hand 8 ft. Patient Line and the 8 ft. Bifurcated
Patient Line to the sterile field.
2.
Receive both ends of the Patient Return Line
from the sterile field.
3.
Connect LUER end of the Patient Return Line
to the LUER lock on the Heat Exchanger.
Observe directional ARROW imprinted on the
patient line.
4.
Receive the “Y-Connection”
end of the
Bifurcated Outflow Patient Line from the sterile
field and connect one end to top of the
reservoir, marked PATIENT RETURN. The
other end is used to remove waste at the end of
the procedure. MAKE SURE THE ON/OFF
PINCH CLAMP ON THIS LINE IS PINCHED
CLOSED.