Chapter 2: Operation
2-12
MAINTAIN HYPERTHERMIC LAVAGE
Perfuse Screen
Routinely check patient and system parameters, on
screen. Respond to and correct system alarms.
Spike additional sterile crystalloid solution, as needed, per
the surgeon.
MAIN OPERATION SCREEN
RATE
The actual rate pumped.
VOL
The actual volume pumped.
P
The actual in-line pressure.
RATE
▲
Press to increase the flow rate (by 10
ml/min). Press and hold to increase the
rate more rapidly. The maximum rate is
1000 ml/min.
RATE
▼
Press to decrease the flow rate (by 10
ml/min). Press and hold to decrease the
rate more rapidly. The minimum flow rate
is 10 ml/min.
1000
ml/min Press to set the system to pump at
RATE
1000 ml/min.
Tpump
The actual fluid temperature as it exits the
heat exchanger. The desired output
temperature can be set using TARGET
▲
/TARGET
▼
key.
T1
Patient Temperature at location 1.
T2
Patient Temperature at location 2.
T3
Patient Temperature at location 3.
T4
Patient Temperature at location 4.
TARGET
▲
Press to increase the output
temperature through the range of 37°C to
48°C. Increment by 0.1°C. Press and hold
to increase temperature more rapidly.
TARGET
▼
Press to decrease the output
temperature through the range of 37°C to
48°C. Decrement by 0.1°C. Press and hold
to decrease temperature more rapidly.
STOP
Temporarily halts pumping and heating.
Status display continues to be active.