Chapter 2: Operation
2-11
INITIATING HYPERTHERMIC LAVAGE
Patient Line Primed and Infuse Screen
T2 = 42.0°C
T4 = 42.0°C
RATE = 1000 ml
min
VOL = 16.2 L
P = 125 mmHg
Tpump = 42.0°C
T1 = 42.3°C
RATE
TARGET
42.5°C
TARGET
42.5°C
1000 ml
min
RATE
STOP
RATE
T3 = 42.0°C
Operation Screen
1.
Press PERFUSE to start infusing at 10 ml/min.
2.
Press 1000 ML/MIN key to pump at 1000
ml/min or adjust flow rate, as needed, by
pressing RATE
▲
/RATE
▼
key.
3.
Adjust the output temperature from 37° to
48°C by pressing TARGET TEMP
▲
/TARGET
TEMP
▼
key to reach the specified output
temperature.
The set temperature is displayed in both
TARGET TEMP
▲
and TARGET TEMP
▼
keys.
The actual fluid temperature as it exits
the heat exchanger, Tpump, is also
displayed on screen.
The
Target
Temperature
is
increased/decreased by 0.1°C every time key
is depressed.
4.
Pump the prescribed solution as directed by
the surgeon. Adjust vacuum to facilitate fluid
return (less vacuum – less return to reservoir).
WHEN PT. LINE PRIM ED PRESS
STOP AND THEN PERFUSE.
STOP
PT. LINE
PRIM E
PERFUSE