Operation BDL
METPOINT® BDL
91
14.5 Current values
Main menu
Current values
Main menu
Current values
A1
14.6 Alarm overview
Main menu
Alarm overview
This is an
alarm-1
for channel A3 and an
alarm-2
for channel A4!
The
current values
view shows the current
measured values of all the connected sensors.
In the event that the adjusted alarm limits are
exceeded or underrun, the respective meas-
ured value flashes yellow (
alarm-1
) or red
(
alarm-2
).
The individual channels can be selected and
the settings viewed and checked, but
no
changes can be implemented here.
Note:
Changes must be effectuated in the
settings
!
In the
alarm overview
,
you can immediately
see whether the alarm is an
alarm-1
or an
alarm-2
.
This is also evident in other menu items:
Main menu
Current values
and in
Main
menu
Settings
Sensor setting
The channel indication flashes yellow for an
alarm-1
an
d red for an
alarm-2
.
Moreover, one can see which relays were set
for which channel as an
alarm-1
and/or as an
alarm-2
.
This is indicated by the yellow and red or
red/yellow squares at the intersections be-
tween measuring channel and relay.
Содержание METPOINT BDL portable
Страница 1: ...EN English Instructions for installation and operation Graphic display recorder METPOINT BDL 10 045 ...
Страница 12: ...Technical data BDL 12 METPOINT BDL 8 3 Housing dimensions ...
Страница 31: ...Connection of the sensors METPOINT BDL 31 11 8 Assignment RS485 Sensor with an RS485 interface ...
Страница 88: ...Operation BDL 88 METPOINT BDL Main menu Graphics ...
Страница 105: ...METPOINT BDL 105 ...
Страница 106: ...106 METPOINT BDL ...
Страница 107: ...METPOINT BDL 107 ...