background image

6.5.2  Low input & display calibration  'In-LO'

This function defines the lower voltage input and
the corresponding transmitter display.   See
example below:
 
         

   Input      Transmitter

             mV           Display

-10.00

-3.000

Low input / display

12.55 

3.765

High input / display

The calibration is performed using the internal
references and may be made with any transmitter
input voltage.

Select 'In-LO' from the menu and press 

P

 which

will access a sub-menu containing two functions
'InPut' and ''dISP'.   

Select the input voltage 'InPut' and press 

P

 which

will cause the specified lower input millivoltage to
be displayed in the form XX.XX The flashing digit
of the input voltage can be changed by pressing
the 

Up

 or 

Down

 buttons.  When the first digit is

correct, pressing 

P

 will transfer control to the next

digit.  When the least significant digit has been
adjusted, press 

E

 to return to the 'InPut' prompt,

from which the display prompt 'dISP' can be
selected by pressing the 

Up

 or 

Down

 button.  

Pressing 

P

 will reveal the transmitter display

corresponding to the specified low input voltage.
The position of the dummy decimal point will be as
 defined by function 'd.P.' - see section 6.5.1

The flashing digit of the display can be changed by
pressing the 

Up

 or 

Down

 buttons.  When the first

digit is correct, pressing 

P

 will transfer control to

the next digit.  When the least significant digit has
been adjusted, press 

E

 to return to the ''dISP'

prompt. 

6.5.3  High input & display calibration  'In-HI'

This function defines the higher voltage input and
the corresponding transmitter display.   See
example below:

         

   Input      Transmitter

             mV           Display

-10.00

-3.000

Low input / display

12.55 

3.765

High input / display

The calibration is performed using the internal
references and may be made with any transmitter
input voltage.

Select 'In-HI' from the menu and press 

P

 which will

access a sub-menu containing two functions 'InPut'
and ''dISP'.   

Select the input voltage 'InPut' and press 

P

 which

will cause the specified higher input millivoltage to

be displayed in the form XX.XX The flashing digit
of the input voltage can be changed by pressing
the 

Up

 or 

Down

 buttons.  When the first digit is

correct, pressing 

P

 will transfer control to the next

digit.  When the least significant digit has been
adjusted, press 

E

 to return to the 'InPut' prompt,

from which the display prompt 'dISP' can be
selected by pressing the 

Up

 or 

Down

 button.  

Pressing 

P

 will reveal the transmitter display

corresponding to the specified higher input voltage.
The position of the dummy decimal point will be as
 defined by function 'd.P.' - see section 6.5.1

The flashing digit of the display can be changed by
pressing the 

Up

 or 

Down

 buttons.  When the first

digit is correct, pressing 

P

 will transfer control to

the next digit.  When the least significant digit has
been adjusted, press 

E

 to return to the ''dISP'

prompt. 

6.6  4/20mA output current calibration 'CAL'

This sub-menu defines the relationship between
the transmitter display and the 4/20mA output
current.  The transmitter output current is derived
from the transmitter display and will change if the
display calibration is changed.     'ZEro' defines the
display at which the output is 4mA, and 'SPAn' the
display at which the output is 20mA.   

To calibrate the output current select 'CAL' from
the programme menu and press 

P

 which will

access a sub-menu containing two functions.
Select 'ZEro' and press 

P

 which will show the

transmitter display at which the output current is
4mA.  The flashing digit of the display can be
changed by pressing the 

Up

 or 

Down

 buttons.

When the first digit is correct, pressing 

P

 will

transfer control to the next digit.  When the least
significant digit has been adjusted, press 

E

 to

return to the 'ZEro' prompt, from which the span
prompt 'SPAn' can be selected by pressing the 

Up

or 

Down

 button.  

Press 

P

  to show the transmitter display at which

the output current is 20mA.  The flashing digit of
the display can be changed by pressing the 

Up

 or

Down

 buttons.  When the first digit is correct,

pressing 

P

 will transfer control to the next digit.

When the least significant digit has been adjusted,
press 

E

 to return to the 'SPAn' prompt. 

Note:

 If a complete loop including a primary

element, BA374C temperature transmitter and a
load are being calibrated, the Trim Menu is the
preferred method. It allows the transmitter display
and output current to be adjusted to compensate
for inaccuracies in the other components in the
loop.  Details of the trim menu are contained in
section 7 which also describes periodic calibration.

      15

Содержание BA374C

Страница 1: ...BA374C Intrinsically safe field mounting indicating temperature transmitter Issue 5 3rd December 2003...

Страница 2: ...ut current calibration 6 7 Security code 6 8 Calibrating the internal references 6 8 1 Conditioning internal references 6 8 2 Voltage input reference 6 8 3 3 wire RTD reference 6 8 4 4 wire RTD refere...

Страница 3: ...Section 6 you will find information on setting up for Section 6 3 Thermocouple input Section 6 4 Resistance thermometer input Section 6 5 Voltage input A word of warning Section 6 8 explains how you...

Страница 4: ...4...

Страница 5: ...ing look up tables stored in permanent memory The resulting temperature is shown on the internal display and controls the 4 20mA output current The transmitter and the optional alarms are powered from...

Страница 6: ...BA374C to be installed in all Zones and to be used with most common industrial gases at ambient temperatures between 40 60o C Note minimum operating temperature 20o C 3 4 Certification information la...

Страница 7: ...Any certified diode return barrier may be used in the return wire For guidance System Certificate Ex96D2505 list some of these devices Fig 3 BA374C powered from a common supply When designing any loo...

Страница 8: ...d such that the environmental conditions specified by the instru ment datasheet and the EC Type Examination Certificate are not exceeded Please consult BEKA associates if the instrument is to be insta...

Страница 9: ...ogramme menu calibration procedure using the trim menu and the calibration example in sections 6 1 7 1 and 8 are read prior to changing function or calibrating the transmitter Figs 7A 7B and 9 show th...

Страница 10: ...ctions in the programme menu is described in the following summary with a cross reference to more detailed information Figs 7A and 7B illustrates the location of each function within the menu Display...

Страница 11: ...11...

Страница 12: ...12...

Страница 13: ...will reveal the current setting This setting may be changed by scrolling through the menu using the Up or Down buttons When the required input is displayed pressing P will reveal a sub menu enabling...

Страница 14: ...esistance thermometer input from the InPut sub menu press P to reveal the tYPE sub menu Press P again to display the current type of resistance thermometer which may be changed by scrolling through th...

Страница 15: ...transfer control to the next digit When the least significant digit has been adjusted press E to return to the InPut prompt from which the display prompt dISP can be selected by pressing the Up or Dow...

Страница 16: ...Voltage input 3 wire RTD 4 wire RTD Current output The input voltage reference must be adjusted first irrespective of the transmitter application but it is only necessary to adjust the other three ref...

Страница 17: ...ly 4mA The transmitter output current shown on the external current measuring instrument should be adjusted to 4 000 0 001mA using the transmitter Up and Down buttons When correct press the E push but...

Страница 18: ...temperature of the primary element operate the Up or Down buttons until the transmitter display is correct Immediately the Up or Down button is operated the annunicator in the top left hand corner of...

Страница 19: ...ple With tHC displayed press P to enter the thermocouple sub menu The first parameter tYPE will be displayed pressing P again will give access to the seven different types of thermocouple Select J usi...

Страница 20: ...erature although this may be less accurate With the transmitter displaying the ZEro prompt and the correct input connected to the transmitter push P which will cause the transmitter to display the the...

Страница 21: ...hazardous area but only certified test equipment should be used unless a gas clearance certificate is available If a BA374C fails after it has been functioning correctly the procedure shown in sectio...

Страница 22: ...74C can be supplied with two solid state single pole alarm outputs Each output may be programmed as a high or low alarm with a normally open or a normally closed output Fig 10 illustrates the conditio...

Страница 23: ...ary of programmable alarm functions Display Summary of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 11 3 1 SP1 Alarm setpoint 1 A...

Страница 24: ...24 Fig 13 Programme structure for optional alarms...

Страница 25: ...rm output status select no nc from the alarm sub menu and press P reveal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return the no nc...

Страница 26: ...AcSP prompt in the Programme Menu Code 0000 will disable the security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments are supplied with thi...

Страница 27: ...or galvanic isolator may be used providing the output parameters do not exceed Uo 30V dc Io 159mA Po 0 8W For guidance the system certificates list some of the devices which may be used Fig 15 Backli...

Отзывы: