66
SUMMARY
Display
Description of function
‘Function’ Function of instrument
Enables the BA364D to be
conditioned as a counter, timer,
tachometer or clock. After making
the selection ‘SurE’ has to be
entered to confirm the instruction
Select ‘tACHO’ to enable the
tachometer functions.
See section 4.5.2
'InPut A'
Type of input
Conditions input A to work with
one of five types of transducer:
‘UOLtS’
Voltage pulse
‘COIL’
Magnetic pick-off
‘Pr.dEt’
NAMUR proximity
detector.
‘COntACt’ Switch contact
‘OP.COL’ Open collector
See section 4.5.3
'd.P.'
Position of decimal point
Defines the position of the decimal
point in speed display.
Note:
The run-time display has a
fixed decimal point to give
resolution of 0.1 hours.
See section 4.5.4
‘dISPLAY.2’Lower display
(run-time)
Turns the lower six digit display,
which shows the run-time, on or
off.
See section 4.5.5
‘r SCALE’
Rotational scale factor
Dividing factor that compensates
for the number of pulses the
transducer produces per
revolution.
May be adjusted between 0.001
and 99999999.
See section 4.5.6
't-bASE'
Timebase
Selectable multiplier to display
rotational speed in revolutions per
second, per minute or per hour.
Select:
tb-1
for revs. / second
tb-60
for revs. / minute
tb-3600
for revs. / hour
See section 4.5.7
Display
Description of function
'FILtEr'
Display filter
Adjustable digital filter to reduce
noise on the speed display.
Adjustable between 0 and 99.
See section 4.5.8
‘CLIP-OFF’ Clip-off
Sets the displayed speed threshold
below which the run-time display
does not operate.
See section 4.5.9
‘LOC-rSEt’ Local reset of run-time display
When turned ‘On’ the run-time
display may be reset to zero from
the display mode by
simultaneously operating the
Up
and
Down
push-buttons for at least
two seconds.
See section 4.5.10
‘CLr.Gtot’
Clear grand total
This function resets the run-time
grand total to zero when ‘CLr YES’
is selected, and ‘SurE’ is entered
to confirm the instruction.
Note:
Once cleared, a grand total
can not be recovered.
See section 4.5.11
'COdE'
Security code
Defines a four digit numeric code
which must be entered to gain
access to the programmable
functions. Default code 0000
disables the security function and
allows unrestricted access to all
programmable functions.
See section 4.5.12