background image

www.bedfont.com

11

NOxBOX

®

 mobile Operating Manual

English

Initial Purge

IMPORTANT: Residual NO gas left in the 

system may convert to NO

2

 during storage. This 

purge procedure MUST be followed prior to use 

on a patient to ensure that any residual gas from 

previous use has been flushed out. 

SAFETY: Refer to all safety notices regarding 

handling the gas and using suitable ventilation 

to prevent build-up of unsafe atmospheres. 

During purge, the exhaust from the unit should 

be fed through an appropriate atmospheric 

scrubber or a scrubber filter for NOx gases. 

This will safely remove the toxic nitrogen oxides 

from the gas exhaust. Contact Bedfont or your 

distributor for NOx scrubber information. 

NOTICE: If system is left pressurised following 

purge for more than 4 hours without use, this 

purge procedure must be repeated to ensure 

safe application.

 

● Ensure the Changeover Dial (See Fig.5, 

item B) arrow is pointing to cylinder 1. Turn 
on cylinder 1 by opening the cylinder valve 
carefully. 

 

● The Pressure 1 gauge dial should register 

the contents pressure of the cylinder (See 
Fig. 5, item A). Ensure there is sufficient 
pressure in the cylinder to commence 
treatment. A full cylinder is approximately 
150 - 200 bar (check with your supplier). A 
cylinder is “empty” at 10 bar.

 

● The outlet pressure gauge dial (Fig. 5, item 

E) should read 2 bar. 

 

● Open the valve on the flow meter (Fig. 5, 

item G) until the flow reads approximately 
600 cc/min. IMPORTANT: do not open valve 
beyond maximum flow indicated on the flow 
meter (Fig. 3, item H) as this may damage 
the instrument control.

 

● Allow the system to purge for 10 minutes 

before switching off cylinder 1 at the cylinder 
valve. Do not over-tighten the cylinder valve 
when closing.

 

● Repeat this purge procedure for cylinder 2. 

 

● Close the valve on cylinder 2 and carefully 

close-off the flow meter. NOTE: when fully 
closing the valve, be careful to not over-
tighten as this may cause damage to the 
control valve mechanism.

 

● Ensure both cylinder valves are shut off 

by end of the purge procedure and prior to 
connecting to patient.

Figure 5

Содержание NOxBOX mobile

Страница 1: ...NOxBOX mobile breath analysis is the new blood test www bedfont com NOxBOX mobile with NOxBOX or NOxBOX O2 ...

Страница 2: ......

Страница 3: ...System Contents 5 6 Instrument Layout 5 7 General Description 6 8 Manual Handling 7 9 Maintenance 8 10 Commissioning 8 11 Installing a Gas Cylinder 9 12 Cylinder Changeover 10 13 Initial Purge 11 14 Connecting to a Patient Circuit 12 13 15 Decommissioning 14 16 Dose Calculation 15 17 Manual Bagging 16 18 Calibration 17 19 Spares 18 20 Specification 18 21 Warranty 19 22 Troubleshooting 20 21 ...

Страница 4: ...nce and safety instructions included in this manual and any manuals supplied with the INO Therapy monitor being used Only trained and responsible personnel should work with or around this equipment To avoid serious injury pay attention to all precautionary labels attached to equipment cylinders containers and boxes prior to start up Do not remove or obscure any label If a label is missing becomes ...

Страница 5: ...th atmospheric oxygen to form NO2 Nitrogen Dioxide NO2 NO reacts with atmospheric oxygen to produce NO2 the NO2 may react further with water to produce nitric and nitrous acid which can both be extremely corrosive The human lungs are a wet environment so inhalation of this extremely toxic gas can have severe effects and every reasonable measure should be made to minimise exposure to NO2 Current UK...

Страница 6: ... cylinder pressure gauges B Changeover unit C NOxAIR NO2 environmental sensor D Single stage regulator E Outlet pressure gauge F Trolley G Low flow indicator H High flow indicator I NOxAIR NO environmental sensor J Spanner attached by chain System Contents E F G H I J A B C D A NOxBOXO2 monitoring instrument 1 NOxBOXO2 2 Water trap Also supplied with Mains power adapter Calibration screwdriver Bac...

Страница 7: ...safe use and maintenance Environmental Alarms Optional NOxAIR environmental gas detectors can be provided to warn personnel of any leaks in the NO supply These portable alarms continuously monitor the atmosphere for NO or NO2 levels depending on model capable of measuring down to 0 1ppm Each monitor has a set of alarms that can be set to ensure best practice environmental monitoring These alarms c...

Страница 8: ...lines or gas cylinders When in use and attached to the patient the NOxBOXmobile should be placed to avoid any tension within the connection lines to the patient circuit The system brakes should be applied to prevent accidental movement of the equipment which could cause lines to disconnect from the patient circuit or damage to any monitor connections Gas cylinders can be heavy Appropriate manual h...

Страница 9: ...s If any evidence of contamination is noted clean the part again Do not immerse the system in cleaning fluids Do not use abrasive cleaning materials on the system as this may damage the protective finish or affect the legibility of control panel items NOTE If using with a NOxBOXO2 or NOxBOX monitor ensure the guidelines for cleaning these monitors are followed separately These monitors can be easi...

Страница 10: ...platform Prior to connecting the cylinder Ensure the gas control valve is shut this should be the default position for the valve Any leaking unregulated gas should be very obvious Remove any safety caps from the threaded valve connection on the cylinder Cylinder connections vary depending on territory Refer to your local distributor for details on your connection type Figure 2 BS14 connector with ...

Страница 11: ... of any connections in place to other systems e g patient circuit before moving the system to avoid stretching lines 2 Turn the auto changeover control so that the arrow is pointing at the new full supply cylinder i e away from the cylinder about to be replaced see Fig 4 where cylinder 2 is being changed This step is best practice if the control remains pointing at the empty cylinder when the supp...

Страница 12: ...er 1 Turn on cylinder 1 by opening the cylinder valve carefully The Pressure 1 gauge dial should register the contents pressure of the cylinder See Fig 5 item A Ensure there is sufficient pressure in the cylinder to commence treatment A full cylinder is approximately 150 200 bar check with your supplier A cylinder is empty at 10 bar The outlet pressure gauge dial Fig 5 item E should read 2 bar Ope...

Страница 13: ...6 Refer to the NOxBOXO2 monitor User Manual for correct set up of the monitor to patient circuit 4 The system is now ready to commence delivery of NO to the patient 5 Refer to the NOxBOXmobile Quick StartGuide QSG for dose setting tables An illustrated example is shown in Figure 6 for 1000ppm supply patient vent flow of 6L min desired dose of 15ppm the NO flow setting should be approx 90 cc min 6 ...

Страница 14: ...ting required to achieve the desired dose Variations between ventilator set ups and individual systems may require slight adjustments e g 10 to this guideline flow rate to achieve the exact dose level If large adjustments are required to the guideline flow rate to achieve the dose rule out any system error with the equipment monitoring or delivery systems by following the troubleshooting actions i...

Страница 15: ... materials 3 Connect the outlet of the NOxBOXmobile to a NOx scrubber filter or an appropriate atmospheric scavenger Contact Bedfont or your distributor for details of the NOx scrubber 4 Open the flow meter control until the NO flow reading is approximately 600 cc min Observe the inlet high pressure gauges Fig 8 A these will gradually decrease until the high pressure system is empty Then observe t...

Страница 16: ...d patient dose of 4ppm the set NO flow cc min NO flow cc min 1000 x 5 5000 20 cc min 1000 4 1 249 NOTE the actual patient dose being delivered should always be confirmed by use of a continuous INO monitor designed for the purpose such as the NOxBOXO2 The guideline NO flow calculated using this formula should be considered as a guide value Fine tune the actual delivered dose by making an appropriat...

Страница 17: ... regular patient ventilator circuit via the dose line exiting the left of the unit need to label this up Fig 9 This inserts into the circuit in the same location described within the Connect to Patient section page 12 The NO flow will be delivered normally through this dose line to the patient ventilator circuit when the control lever is turned such that the arrow points towards this dose line Ven...

Страница 18: ...s equipment can only be undertaken by the original manufacturer These are safety critical devices and must not be tampered with as this could put patients at risk Pressure Regulators The pressure regulator performance should be checked and the output pressure verified as 2 bar All calibration repair or refurbishment of this equipment can only be undertaken by the original manufacturer These are sa...

Страница 19: ...sensor for NOxBOXO2 NOXO2 NO2 Replacement NO2 sensor for NOxBOXO2 NOXO2 O2 Replacement O2 sensor for NOxBOXO2 NOXO2 NAF Nafion drying tube NOXO2WT COMPLETE Water trap FIX053 Power lead UK FIX054 Power lead EU NOX Y CAL Calibration system consisting of Y piece and male luer NOXKIT V Selection of connectors NO and O2 to vent NOXO2 FIL Bacterial filter CON139 Leur fitting for NOxBOXO2 CON136 Leur fit...

Страница 20: ... of five years from the date of shipment Bedfont s sole obligation under this warranty is limited to repairing or replacing at its choice any item covered under this warranty when such item is returned intact prepaid to the factory The warranty does not apply to any products which have been repaired or altered by unauthorised persons or which have been subject to misuse negligence or accident ...

Страница 21: ...ered to the patient WARNING sudden withdrawal from NO therapy can have severe effects on the patient including but not limited to hypoxic rebound Ensure that best practice is observed at all times to ensure the patient is not put at unnecessary risk If the above procedure cannot be followed for a quick system purge then the system must be purged through the flow meters This must be done without co...

Страница 22: ... rises as expected but then there is a noticeable drop before the float rises again Why is this A The fine control valve on the flow meter may have been damaged by excessive opening of the valve control This can create deep scoring on the valve surface which affects the flow control performance across the flow range Please report to Bedfont or your distributor for corrective action Q Do Bedfont su...

Страница 23: ......

Страница 24: ...te this literature without prior notice Registered office England and Wales Registered No 1289798 Station Road Harrietsham Maidstone Kent ME17 1JA England Tel 44 0 1622 851122 Fax 44 0 1622 854860 Email ask bedfont com www bedfont com Bedfont Scientific Ltd ISO 9001 2008 Cert No FM 31664 ISO 13485 2003 Cert No MD 502905 ...

Отзывы: