2-2
Learning to Create Methods for the Biomek® 3000
Beckman Coulter, Inc.
2.2
Creating a New Method
In this chapter, you will build a simple method that transfers liquid between labware.
In building this method, you will:
•
Populate the deck with labware using the
Instrument Setup
step.
•
Transfer liquid with an MP200 Multi-Tip Pipette Tool using a
Transfer
step.
•
Move labware with the Gripper Tool using a
Move Labware
step.
Note:
A gripper tool rack must be installed on the Biomek 3000 deck to
store a gripper tool when not in use.
When you create a new method by dragging a step into the Method View, the
Start
and
Finish
steps appear automatically. You will use an
Instrument
Setup
Step to tell
Biomek Software where you have physically placed labware on your deck.
Before proceeding with this quick-start method, make sure that:
•
an
MP200
Multi-Tip Pipette Tool and
Gripper
Tool are added in
Hardware
Setup
•
the deck layout in the
Deck Editor
contains at least one
GripperToolRack
,
one
ManualLatch
, and four
Biomek3000Positions
Note:
If your instrument configuration does not have a gripper tool, a
standard
ToolRack
position can be used instead of the
GripperToolRack
.
•
the deck is framed (refer to the
Biomek® 3000 Laboratory Automation
Workstation User’s Manual,
Chapter 13,
Framing the Biomek® 3000
).
If using Accounts &
Permissions . . .
You must have
Develop
Methods
permission to
create methods. Refer to
the
, for more
information.